7-13
7 Technical Specifications
Signal Level Depending on Surface and Stand-Off Distance
Figure 7.3:
Typical signal level range depending on the stand-off distance measured on a diffuse white surface
and 3M Scotchlite™ Tape (retro-reflective film)
Table of the Visibility Maxima
7.5.3
Scanner
Visibility maxima (mm) for l = 204 mm
141
345
549
753
957
1161
1 365
1 569
1 773
1 977
2 181
2 385
2 589
2 793
2 997
3 201
3 405
3 609
3 813
4 017
4 221
4 425
4 629
4 833
5 037
Add for every additional visibility maximum 204 mm.
Type:
Servo-controlled Galvo scanner
Maximum deflection:
Horizontal: ± 25°
Vertical: ± 20°
Angular resolution:
16-Bit
Point stability:
< 0.001° / hour (after 2 h warm-up period, with
constant ambient temperature)