Set speed
You can set the adjustment speed of the telescope drive. To do so, proceed
as follows:
1. Reduce the adjustment speed by turning the potentiometer on the hand
set clockwise (to the right).
Raise the adjustment speed by turning the potentiometer on the hand set
counterclockwise (to the left).
2. Lock the set speed with the lever on the potentiometer.
A.5.4.5 Technical Specifications
Drive Control
Telescope Drive
Mains voltage:
230 V / 50 Hz or 115 V / 50-60 Hz
Protection rating:
IP 30
Dimensions:
334 mm x 62 mm x 110 mm
Weight:
2.6 kg
Protection rating:
IP 30
Operating temperature:
+ 5°C … + 40°C (41°F … 104°F)
Operating time:
20% (in 10 min. length)
Velocity stages:
stageless
Max. velocity:
28 mm / s
Max. travel:
350 mm
Dimensions:
refer to
A.12 and
Weight:
27 kg (vertical test stand)
30 kg (horizontal test stand)