76
20
μm to 50μm. The actuator is plugged into channel 2 (amplifier 3) --> position 3 f.l.. It should stay
on 20
μm for 50ms and on 50μm for 150ms. The value from measurement system is then sent to
the monitor output.
commands:
- cl,2,1
Enter
closed loop amplifier 3 on
- gfkt,2,3
Enter
generator function rectangle
- gfrec,2,5
Enter
frequency rectangle = 1 / (50ms+150ms) = 5Hz
-
garec,2,37.5
Enter
amplitude rectangle = (50μm – 20μm) / 80μm * 100% = 37.5%
- gorec,2,25
Enter
offset rectangle = 20
μm / 80μm * 100 % = 25 %
- gsrec,2,25
Enter
symmetry rectangle = 50ms / (50ms + 150ms) * 100% = 25%
- monsrc,2,0
Enter
set monitor source to position value in closed loop