CEO: Dr. Bernt Götz • phone +493641/66880 • fax +493641/668866 • www.piezosystem.com
61
garun
Switching on/off the AutoRun of the
generators
garun,channel,value
0 = switch off
1 = switch on
r/w
trgss
trigger generation stroke position
start
minimum: >0.2% of total stoke to
maximum: total stroke minus 0.2% of
total stroke [micron] or [mrad]
r/w
trgse
trigger generation stroke position
end
minimum: >0.2% of total stoke to
maximum: total stroke minus 0.2% of
total stroke [micron] or [mrad],
always keep: trgse > trgss !
r/w
trgsi
trigger generation position intervals
in % CL-stroke
>0.05% of total stroke in closed loop
[micron] or [mrad]
r/w
trglen
duration of trigger impulses
n*20µs
n=1 to 255
r/w
trgedge
trigger generation edge
0= trigger off
1= trigger at rising edge
2= trigger at falling edge
3= trigger at both edges
4= reversal motion
5= inverted as 4
7= length of impulse as 4
trglen
r/w
recstart
start data recorder
w
recstop
stop data recorder
w
recast
Autostart data recorder
starts with the next set-command
0
– off
default: 0
1
– on
r/w
recsrc3
data recorder data source
recsrc,source1,source2,source3
see 10.5 Datenrecorder
r/w
recstr
stride / step width
recording all recstr *20µs
1 to 2
32
-2
r/w
recwridx
current write index
to determine the current write position
after the manual stop
r
recrdidx3 set selection index
0 to reclen
r/w
recrd
reads incrementally the buffer out
channel 0 to 3,optionally: number of
the values 1 to reclen
r
reclen
length of the records in samples
1 to 500000 samples per channel
r/w
table 3: commands channel depended