User Manual
E727T0005, valid for E-727
BRO, 2019-06-28
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Page 61 / 240
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
Communication via the RS-232 Interface
The baud rate of E-727 for RS-232 communication is preset as follows:
Uart baud rate (parameter ID 0x11000400): 115200
Further possible values are 9600, 19200, 38400, 57600.
To successfully establish communication, the baud rates of the E-727 and PC must match.
Establishing communication via RS-232
The procedure for PIMikroMove is described in the following. The procedure for other PC software
programs (PITerminal, drivers for use with NI LabVIEW software) is similar.
1.
Start PIMikroMove.
The
Start up controller
window opens with the
Connect controller
step.
−
If the
Start up controller
window does not automatically open, select the
Connections > New
...
menu item in the main window.
2.
Select
E-727
in the field for controller selection.
3.
Select the
RS-232
tab on the right side of the window.
4.
In the
COM Port
field, select the COM port of the PC to which you have connected the E-727.
5.
Set a suitable value in the
Baudrate
field to adapt the baud rate of the PC to the baud rate of
the E-727.
6.
Click
Connect
to establish communication.
If communication has been successfully established, the
Start up controller
window switches
to the
Start up axes
step.