Parameter
ID (hexa-
decimal)
Data
Type
Command
Level for
Write
Access
Item Type
Parameter
Name
Description
0x0e000e00 INT
1
System
Adv. Piezo
Control License
Parameters of the Advanced
Piezo Control option (p. 19). See
the separate E712T0007
Technical Note for more
information.
0x0e000f00
INT
3
System
Adv. Piezo
Control License-
Valid Flag
0x0E001500 INT
1
System
Reboot on DIO
Input
The value of the parameter
enables/disables the Reset input
on pin 2 of the MDR14 socket
(details see p. 227).
0x0E001D00 INT
1
System
Trigger Input
Filter Enable
The value of the parameter
enables/disables a filter for the
digital trigger input as follows:
0 = OFF: Filter is disabled.
Processing of the trigger input is
done immediately.
1 = ON: Filter is enabled to
reduce the influence of noise at
the trigger input line. Default
setting.
0x0E001E00 INT
1
System
Discon. Target
Man. In With
Stop
Determines the behaviour when
the analog input is used as a
control source. See „Using the
Analog Input“ (p. 85) for details.
0x0e001f00
INT
1
System
Check Sensor
Position
Plausibility
For future use.
Enables oder disables plausibility
check of sensor position.
0x0e002000 INT
1
System
Move To Last
Commanded
Position
Determines the target positions
to be set when switching on the
servo mode:
0 = The target positions are set to
the current positions of the axes
(default).
1= The target positions are set to
the last commanded target
positions for the axes. This can
cause an abrupt motion.
0x0e002400 INT
1
System
Trigger Input
Filter Samples
0x0f000000
INT
1
System
Power Up Read
ID-Chip
For details, see “ID-Chip Support /
Stage Replacement” (p. 131).