rise time and settling time of the axis, based on the recorded step response
measurement.
d)
Check the step response result and compare it with the examples shown in the
figures below.
The rise rate of the step response is very low in the figure below. This means that
the P term is too low and has to be increased.
The figure below shows a step response with a high overshoot which means that
the P term is too high and has to be decreased.