Note that depending on the selection for
Automatic I-Term calculation
, the
Servo-
Loop I-Term
value is changed too automatically when you change the
Notch
Frequency 1
value (for details, see p. 134).
b)
If you have measured a second resonance peak, enter the frequency value of the
second resonance peak in the
Notch Frequency 2
field (in Hz). If the second notch
filter is deactivated (rejection value = 1), change the rejection value to 0.05 in the
Notch Rejection 2
field.
8.
Save or discard the new settings in the
Parameter Settings
panel:
−
If you want to keep the new settings, save them to the nonvolatile memory of the
E-727 by clicking the
Save as Default (EEPROM)
button.
−
If you want to discard the new settings and reset the parameter values to their
defaults (i.e. to their values from nonvolatile memory), click the
Reset All to
Default
button.