6 Patient Alarms and INOPs
77
CHECK TI CONFIG
Monitor
If this INOP appears, the monitor can be used normally, except
for the Trace Interpretation feature. Contact your service
personnel.
Check Touch Input
Monitor
Perform a visual and functional check of the touch input device.
Contact your service personnel.
COINCIDENCE
Monitor
One or more fetal/maternal heart rates/pulse rates persistently
coincide with each other (see “Cross-Channel Verification” on
page 97).
CUFF NOT DEFLATED
Numeric is replaced by a -?-
INOP tone.
During this INOP, alarms
cannot be paused or switched
off.
Noninvasive
blood pressure
Remove the cuff from the patient. Make sure that the tubing is
not kinked or twisted. Try restarting the measurement. You can
silence the INOP, but the INOP message remains visible until
the next measurement is started or the
Stop All
SmartKey is
selected.
DFHR1 EQUIP MALF
DFHR2 EQUIP MALF
DFHR3 EQUIP MALF
INOP tone
DECG
There is a problem with the DECG hardware. Contact your
service personnel.
DFHR1 LEADS OFF
DFHR2 LEADS OFF
DFHR3 LEADS OFF
Numeric is replaced by a -?-
INOP tone.
DECG
One or more DECG lead is not attached. Make sure that all
required leads are attached, and no electrodes have been
displaced. Check all connections are sound, and that the legplate
attachment electrode is properly attached. If the INOP persists,
try using another adapter cable, or legplate attachment electrode.
If the INOP still persists, contact your service personnel.
DFHR1 SIGNAL LOSS
DFHR2 SIGNAL LOSS
DFHR3 SIGNAL LOSS
DECG
The input signal quality is not sufficient to process the
measurement. Reapply the fetal scalp electrode.
DFHR1 UNPLUGGED
DFHR2 UNPLUGGED
DFHR3 UNPLUGGED
INOP tone
DECG
Reconnect the DECG transducer to the monitor. Check all
connections are sound.
MECG EQUIP MALF
INOP tone
MECG
There is a problem with the MECG hardware. Contact your
service personnel.
MECG LEADS OFF
Numeric is replaced by a -?-
INOP tone.
MECG
One or more MECG lead is not attached. Make sure that all
required leads are attached, and no electrodes have been
displaced. Check all connections are sound. If the INOP
persists, try using another adapter cable. If the INOP still
persists, contact your service personnel.
MECG UNPLUGGED
INOP tone
MECG
Reconnect the MECG transducer to the monitor. Check all
connections are sound.
FetRec EQUIP MALF
INOP tone
Recorder
There is a problem with the fetal recorder hardware. Contact
your service personnel.
INOP Message, Indication
Source
What to do
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...