26 Specifications and Standards Compliance
210
Table 3 - Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Conducted RF Immunity Test EN/IEC 61000-4-6
IEC 60601-1-2 Test Level
over 150 kHz to 80 MHz
Compliance Level
Electromagnetic Environment
Guidance:
Recommended Separation Distance
(
d
)
(in Meters, at Frequency Range
Tested)
for Ultrasound and ECG
Measurements
3.0 V
RMS
3.0 V
RMS
d = 1, 2
√
P
Key:
d
= Recommended separation distance in meters (m)
P
= maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer
V1
= Tested compliance level (in Volts) for the Conducted RF Immunity test IEC 61000-4-6
Note: The device meets the compliance level of 3.0 V
RMS
according to IEC 60601-1-2 over the
specified test frequency range. Over the frequency range 150 kHz to 80 MHz, the recommended
separation distance in meters (d) is found by the following equation:
For a compliance level of 3.0 V
RMS
: d = 1, 2
√
P
Table 4 - Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Radiated RF Immunity Test EN/IEC 61000-4-3
IEC 60601-1-2 Test Level
over 80 MHz to 2.5 GHz
Compliance Level
Electromagnetic Environment
Guidance:
Recommended Separation
Distance (
d
)
(in Meters, at Frequency Range
Tested)
for Ultrasound and ECG
Measurements
3.0 V/m
3.0 V/m
Over 80 MHz to 800 MHz:d =
1, 2
√
P
Over 800 MHz to 2.5 GHz:d =
2, 3
√
P
Key:
d
= Recommended separation distance in meters (m)
P
= maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer
E1
= Tested compliance level (in Volts/meter) for the Radiated RF Immunity test IEC 61000-4-3
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...