26 Specifications and Standards Compliance
200
SpO
2
Alarm Specifications
Range
Adjustment
Delay
SpO
2
50 to 100%
1% steps
(0, 1, 2, 3,... 30) +
4 seconds
Desat
50 to Low alarm limit
1% steps
Pulse
30 to 300 bpm
1 bpm steps (30 to 40 bpm)
5 bpm steps (40 to 300 bpm)
max. 14 seconds
Tachycardia
Difference to high limit 0 to 50
bpm
5 bpm steps
max. 14 seconds
Clamping at 150 to 300 bpm
5 bpm steps
Bradycardia
Difference to low limit 0 to 50 bpm 5 bpm steps
max. 14 seconds
Clamping at 30 to 100 bpm
5 bpm steps
SpO
2
Factory Default Settings
Desat Alarm Limit
80
Low Alarm Limit
90
High Alarm Limit
100
Desat Alarm Limit Delay
20 seconds
Low Alarm Limit Delay
10 seconds
High Alarm Limit Delay
10 seconds
Averaging Time
10 seconds
NBP Alarm Suppression
On
Parameter Alarms On/Off
On
Color
Cyan
Pulse Settings
Pulse Alarms On/Off
On
Pulse High Limit
120 bpm
Pulse Low Limit
50 bpm
Bradycardia: Difference to Low Limit
20 bpm
Bradycardia: Clamp
40 bpm
Tachycardia: Difference to High Limit
20 bpm
Tachycardia: Clamp
200 bpm
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...