26 Specifications and Standards Compliance
194
Auto Offset Correction
3 seconds after connecting the transducer,
the Toco value is set to 20 units
Auto Zero Adjust
Toco value is set to zero following a negative
measurement value for 5 seconds
Maternal Pulse from Toco
Emitted Light Energy
≤
15 mW
Wavelength Range
780 to 1100 nm
Range
40 to 240 bpm
Resolution
1 bpm
Display Update Rate
1 / s
Accuracy
± 2% or 1 bpm, whichever is greater
Update Rate
every 4 seconds
IUP
Measurement Method
Passive Resistive Strain Gauge Elements
Measurement Range
-100 to +300 mmHg
Signal Range
-99 to 127 mmHg
Resolution
Display
1 mmHg
Printer
1/4 mmHg
Sensitivity
5 µV/V/mmHg
Offset Compensation
+100 to -200 mmHg
Accuracy (not including sensor accuracy)
± 0.5% per 100 mmHg
Update Rate
Display
1 / second
Printer
~4 / seconds
Auto Offset Correction
3 seconds after connecting the transducer,
the IUP value is set to 0 mmHg
ECG
Type
DECG
Single Lead ECG (derived from Fetal Scalp
Electrode)
MECG
Single Lead ECG (derived from RA and LA
electrodes)
Measurement Range
30 to 240 bpm
Resolution
Display
1 bpm
Recorder
1/4 bpm
Accuracy
± 1 bpm or 1%, whichever is greater
Beat-to-Beat Change (max.)
MECG: 28 bpm
DECG: 28 bpm (with Artifact Suppression
On)
Differential Input Impedance
> 15M
Ω
Electrode Offset Potential Tolerance
± 400 mV
Performance Specifications
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...