19
155
19
Monitoring Maternal Heart /
Pulse Rate
You can monitor the maternal heart/pulse rate using one of four sources:
•
Maternal heart rate (MHR) via MECG electrodes
•
SpO
2
(pulse rate)
•
Maternal pulse from Toco MP transducer (pulse rate)
•
NBP (pulse rate)
Maternal heart / pulse rates derived from MECG, SpO
2
and Toco MP are
continuous
measurements,
and are compared against the FHR for cross-channel verification. Average pulse rate derived from
noninvasive blood pressure is an
intermittent
measurement, and is therefore not used for cross-
channel verification.
Priority for Maternal Heart / Pulse Rate
Only one Maternal Heart Rate / Pulse Rate numeric will be displayed and recorded at a time (see
priority table). Note that when higher-priority measurements are connected but temporarily not
providing valid numerics, lower-priority numerics may be displayed and recorded instead.
Cross-Channel Verification
It is recommended that you monitor both the maternal pulse rate/heart rate and the fetal heart rates to
reduce the risk of misinterpreting the maternal heart rate for the fetal heart rate. See “Confirm Fetal
Life Before Using the Monitor” on page 10 and “Cross-Channel Verification” on page 97.
SpO
2
pulse rate traces have an averaging calculation of approximately 10 seconds and an overall delay
of approximately 12 seconds (depending on printer speed). This differs from a non-averaged beat-to-
Priority
Maternal Heart / Pulse Rate Source
Used for CCV
Provides QRS Tone
1
HR from MECG measurement
Yes
Yes
2
Pulse from SpO
2
measurement
Yes
Yes
3
Pulse from Toco MP transducer
Yes
No
4
Pulse from NBP measurement
No
No
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...