18 Monitoring SpO2
152
SpO2 Signal Quality Indicator
The SpO
2
numeric is displayed together with a signal quality indicator (if configured and enough space
is available) which gives an indication of the reliability of the displayed values.
The level to which the triangle is filled shows the quality of the signal; the indicator below shows a
medium signal quality, the signal quality is at a maximum when the triangle is completely filled.
Assessing a Suspicious SpO2 Reading
Traditionally, pulse rate from SpO
2
was compared with heart rate from ECG to confirm the validity of
the SpO
2
reading. With newer algorithms, such as FAST-SpO
2
, this is no longer a valid criteria because
the correct calculation of SpO
2
is not directly linked to the correct detection of each pulse.
When pulse rate is very low, or strong arrhythmia is present, the SpO
2
pulse rate may differ from the
heart rate calculated from ECG but this does not indicate an inaccurate SpO
2
value.
WARNING
With pulse oximetry, sensor movement, ambient light (especially strobe lights or flashing lights) or
electromagnetic interference can give unexpected intermittent readings when the sensor is not
attached. Especially bandage-type sensor designs are sensitive to minimal sensor movement that might
occur when the sensor is dangling.
Understanding SpO2 Alarms
This refers to SpO
2
specific alarms. See the Alarms chapter for general alarm information. SpO
2
offers
high and low limit alarms, and a high priority desat alarm. You cannot set the low alarm limit below the
desat alarm limit.
CAUTION
If you measure SpO
2
on a limb that has an inflated noninvasive blood pressure cuff, a non-pulsatile
SpO
2
INOP can occur. If the fetal monitor is configured to suppress this alarm there may be a delay of
up to 60 seconds in indicating a critical status, such as sudden pulse loss or hypoxia.
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...