11 Monitoring Twin FHRs
116
Separating FHR Traces
To help you to interpret traces with similar baselines, you can separate the baselines by an offset of
20 bpm by switching trace separation on. For details of the offset, see “Determining the Separation
Order” on page 116.
Switching Trace Separation On and Off
♦
Connect transducers to the monitor to measure FHR. Depending on the measurement method,
you need
either
two ultrasound transducers
or
FM30/50
One ultrasound and one Toco
+
transducer (to monitor DECG)
♦
Enter the
Main Setup
menu by pressing the
Main Setup
.
♦
Select
Fetal Recorder
.
♦
Select
Trace Separation
to toggle between
On
and
Off
.
♦
Exit the
Main Setup
menu.
Determining the Separation Order
In Configuration Mode, you can choose between two different ways for dealing with the trace offsets
on the recording (the order in which they are separated) when
Trace Separation
is
On
.
1
Enter the
Main Setup
menu by pressing the
Main Setup
SmartKey.
2
Select
Fetal Recorder
.
3
Select
Separation Order
to toggle between
Standard
and
Classic
.
–
Standard
: the FHR2 trace is shifted up by 20 bpm (it is recorded 20 bpm higher than it really
is). No offset is ever applied to the FHR1 trace - it stays where it is. (In case of a third FHR,
this is shifted down by 20 bpm.)
–
Classic
: the FHR1 trace is shifted up by 20 bpm when there is more than one FHR
measurement. No offset is ever applied to the FHR2 trace - it stays where it is. (In case of a
third FHR, this is shifted down by 20 bpm.)
4
Exit the
Main Setup
menu.
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...