8 Non-Stress Test Timer
88
Accessing the NST Setup Pop-up Keys
You control and set up the NST timer (for example, start, stop, or clear the timer, and set the run time)
using a selection of pop-up keys that you access via any one of three possible routes:
•
Via the
Timer
SmartKey (Route 1).
•
Via the
Main Setup
SmartKey (Route 2).
•
Via the NST menu entry in the menu that pops-up when you touch the top left-hand corner of the
screen (Route 3).
Via the Timer SmartKey (Route 1)
Press the
Timer
SmartKey
. The
Timers
window opens, and the pop-up keys for
controlling/setting up the NST timer appear (see “Pop-up Keys for NST Timer Setup” on
page 88).
Via the Main Setup SmartKey (Route 2)
1
Enter the
Main Setup
menu using the SmartKey
.
2
Select
NST
to enter the
Setup NST
menu. At the same time, the pop-up keys for controlling/
setting up the NST timer appear (see “Pop-up Keys for NST Timer Setup” on page 88).
Via the NST Field (Route 3)
Select the NST field displayed in the top left-hand corner of the screen (when so configured). The
pop-up keys for controlling/setting up the NST timer appear (see “Pop-up Keys for NST Timer
Setup” on page 88).
Pop-up Keys for NST Timer Setup
Pop-Up Keys
Selecting this pop-up key lets you....
Comments
Start
start the timer
Stop
stop the timer, allowing either restarting after a pause
(
Start
) or clearing (
Clear
key).
Setup NST
enter the
Setup NST
menu. From here you can set the
run time.
This pop-up key is not available with
Route 2, as the
Setup NST
menu is
already open.
Timer
return to the
Timers
window.
This pop-up key is not available with
Route 1, as the
Timers
window is
already open.
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...