Starting Chromera. 23
Preparing the Calibration Vials for Auto Tuning
There are two calibration vials as shown below. Typically use
Vial 1 for Positive Ion Tune mix
and
Vial 2 for a Negative Ion Mix
.
1.
To remove calibration
Vial 1
, reach in and unscrew the left side calibration vial as shown below.
2.
Remove the vial and fill the vial with approximately 15 mL of tune mix.
3.
Re-insert the calibration vial making sure the small draw tube is inserted into the 50 mL calibration
vial. Also make sure that the calibration vial is securely fastened into the blue cap to avoid leaks
after pressurization.
4.
Connect the peak tubing that extends from the calibration compartment and use a Peak Finger tight
fitting to secure the tubing to the sprayer probe.
Calibration Vials on the SQ 300 MS
Unscrew the calibration vial
The calibration vial removed
Filling the Calibration vial
Vial 1
Vial 2
Summary of Contents for FLEXAR SQ 300 MS
Page 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Page 2: ......
Page 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Page 8: ...6 Flexar SQ 300 MS User s Guide ...
Page 9: ...Starting ...
Page 14: ...12 Flexar SQ 300 MS User s Guide ...
Page 15: ...Starting Chromera ...
Page 36: ...34 Flexar SQ 300 MS User s Guide ...
Page 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Page 55: ...Creating Methods and Sequences ...
Page 68: ...66 Flexar SQ 300 MS User s Guide ...
Page 69: ...Starting Data Acquisition ...
Page 77: ...Analyze Results in Post Run ...
Page 104: ...102 Flexar SQ 300 MS User s Guide ...
Page 105: ...Evaluating Mass Spectra ...
Page 119: ......