Starting Chromera. 29
Set the
source horizontal position
so that the horizontal probe marking is positioned 10
spaces to the right side of the scale as shown below:
Set the
Probe Tilt
position as shown below:
11.
To start auto tune, click on the
Auto Tune
button.
Each Auto Tune takes approximately 20 min.
If the Auto Tune stops abruptly you can re-prime by clicking on the
Sample Prime
button and then
click the
Start Pumping
button and wait an additional 30 seconds before running Auto Tune again.
Set
Horizontal
Set Tilt Position Here