Creating Methods and Sequences. 59
NOTE: This is a quick way to build a Method requiring multiple acquisition experiments. Simply copy previously
defined Tunes as this minimizes the number of changes required for each Tune in the Method.
20.
Click on the top
Sim
blue dot and change it to
Scan
in
Acquisition Function
of the
Global Variables
section.
21.
Set the
Low
m/ z
to 500
and the
High
m/ z
to 625.
22.
Set the
Pulse Counting Time (µs)
to
100.
NOTE: The pulse counting (dwell) time entered here applies to each data point within the scan. In this example,
the scan range is m/z 500-625, and there are 10 samples per mass, resulting in 1251 total data points
acquired per scan, each
23.
Set the
Drying Gas Flow Rate (L/min)
to
12
.
of which will be measured for 100 µs.