Starting Chromera. 33
Running Auto Tune Using the Built-in Syringe Pump
Running Auto tune using the built-in syringe pump is very similar to running Auto Tune using the
Calibration vials except that the syringe parameters and flow rate must be input as demonstrated in the
following screen:
1.
Open
Auto Tune
and select
Syringe
from the
Analyte Source
drop-down menu.
Make sure the syringe is loaded with desired calibrant and that the peek line extends from the
syringe tip and connects to the probe sprayer with a peek finger tight connector.
2.
Click the
Sample Prime
button.
The
Tune Setup
dialog displays.
3.
Type in the
System Name
and the
User Name
.
4.
In the
Syringe Pump
section, select the
Make
of the syringe from the drop-down menu.
When using a 500 uL Hamilton Syringe select
Custom - user entered
in the drop-down and the
set the syringe inner diameter to 3.26 mm and the flow rate to 15 µl/min.
5.
Click the
Start Pumping
button and wait until the mid-range mass shown in the scan window has
stabilized just as you did before when using the calibration vial.
6.
Run
Auto Tune
and
Check Tune
as described earlier in this section.
Summary of Contents for FLEXAR SQ 300 MS
Page 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Page 2: ......
Page 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Page 8: ...6 Flexar SQ 300 MS User s Guide ...
Page 9: ...Starting ...
Page 14: ...12 Flexar SQ 300 MS User s Guide ...
Page 15: ...Starting Chromera ...
Page 36: ...34 Flexar SQ 300 MS User s Guide ...
Page 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Page 55: ...Creating Methods and Sequences ...
Page 68: ...66 Flexar SQ 300 MS User s Guide ...
Page 69: ...Starting Data Acquisition ...
Page 77: ...Analyze Results in Post Run ...
Page 104: ...102 Flexar SQ 300 MS User s Guide ...
Page 105: ...Evaluating Mass Spectra ...
Page 119: ......