Flexar SQ 300 MS User’s Guide
Auto Tune on the SQ 300 MS Detector
Tuning the Flexar SQ 300 MS sets ion source and ion lenses parameters to obtain optimal resolution and
sensitivity, and involves adjusting the probes.
The Flexar SQ 300 MS instrument is tuned at the factory
and it will be checked and re-tuned if necessary during installation. Checking the SQ 300’s performance
by running a Check Tune or Auto Tune is a quick diagnostic procedure if changes in system performance
are observed. This Auto Tune procedure simplifies the tuning process.
Upon successful completion, Auto Tune will create calibrated Tunes at 3 different scan speeds, 1k/sec,
5k/sec and 10k/sec. These Tunes will be the foundation upon which you build an MS acquisition method.
Auto Tune is performed through the SQ 300 MS driver.
About the Samples
PKI Tune Mix
Negative Mode Ions: 92, 205, 531 1166, 1466, 2666
Positive Mode Ions: 123, 195, 506, 1022, 1422, 2222, 2522
Calibrants
There are two calibration mixtures that can be employed at the present time. The PKI positive ion tune
mix which is good for positive ion tuning and the PKI negative ion tune mix for negative ion tuning.
Calibrant Delivery Modes
A calibrant can be introduced into the mass spectrometer by several methods including: using the
calibration vials that are located on the front end of the SQ 300 mass spectrometer, using the on board
syringe pump (if available) which is located on the front right hand side of the SQ 3000 or by using an
external syringe pump.
Summary of Contents for FLEXAR SQ 300 MS
Page 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Page 2: ......
Page 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Page 8: ...6 Flexar SQ 300 MS User s Guide ...
Page 9: ...Starting ...
Page 14: ...12 Flexar SQ 300 MS User s Guide ...
Page 15: ...Starting Chromera ...
Page 36: ...34 Flexar SQ 300 MS User s Guide ...
Page 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Page 55: ...Creating Methods and Sequences ...
Page 68: ...66 Flexar SQ 300 MS User s Guide ...
Page 69: ...Starting Data Acquisition ...
Page 77: ...Analyze Results in Post Run ...
Page 104: ...102 Flexar SQ 300 MS User s Guide ...
Page 105: ...Evaluating Mass Spectra ...
Page 119: ......