52 .
Flexar SQ 300 MS User’s Guide
This indicates that to produce the maximum intensity
m/
z 195 fragment ion, a
Capillary Exit
voltage of ~250 volts should be used.
19.
Select
Ramp
from the
Tune
menu to deselect it.
20.
Stop the analysis, close all
Snapshot screens
, and close the
SQ 300 MS driver
.
Obtaining fragmentation information on a compound can be very valuable for a number of reasons.
It gives insight into the structure of the compound, which is typically applicable to other compounds
with the same base structure (e.g., drugs and their metabolites). It also provides additional ions
associated with the analyte that can be measured instead of, or in addition to, the protonated
molecular ion. For example, in quantitative analyses, there is occasionally an issue with a
contaminant or mobile phase ion at the same nominal mass as the analyte to be measured. This
may have a significant effect on the detection limits of the analyte due to background noise. To
overcome this, fragments of the analyte can be monitored instead of, or in addition to, the
protonated molecular ion. The probability of the interfering ion having a fragment at the same
m/z
value
as a fragment from the analyte is extremely
low (but still must be verified!)
Summary of Contents for FLEXAR SQ 300 MS
Page 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Page 2: ......
Page 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Page 8: ...6 Flexar SQ 300 MS User s Guide ...
Page 9: ...Starting ...
Page 14: ...12 Flexar SQ 300 MS User s Guide ...
Page 15: ...Starting Chromera ...
Page 36: ...34 Flexar SQ 300 MS User s Guide ...
Page 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Page 55: ...Creating Methods and Sequences ...
Page 68: ...66 Flexar SQ 300 MS User s Guide ...
Page 69: ...Starting Data Acquisition ...
Page 77: ...Analyze Results in Post Run ...
Page 104: ...102 Flexar SQ 300 MS User s Guide ...
Page 105: ...Evaluating Mass Spectra ...
Page 119: ......