Analyze Results in Post Run. 91
Creating an Extracted Ion Chromatogram (EIC)
The EIC is a form of a limited mass range TIC. An EIC is extracted from scan data, and typically used
where specific ion(s) of interest, or small mass ranges are being searched for. It is possible to display
several mass ranges in the same EIC chromatogram.
NOTE:
If specific ions of interest are being monitored for quantitative purposes, the best sensitivity (i.e., lowest
detection limits) will be obtained by monitoring the ions in SIM acquisition
An EIC can be generated in two ways:
mode, rather than extracting
them from scan data acquired over a large mass range.
•
Using the EIC dialog.
•
Using a "box" procedure in a generated mass spectrum.
Using the EIC Dialog
To create an Extracted Ion Chromatogram (EIC):
1.
Select the TIC in the SQ 300 MS driver window
2.
Select
EIC
from the
View
menu.
3.
Enter the first mass range in the first row of the
m/ z
Range Selection
table by clicking in the cell
and entering a value.
4.
Enter the second mass range in the second row of the
m/ z
Range Selection
table.
5.
Decide how to display the curves in the
Curve Definition
table.
This allows you to perform simple algebraic operations (addition or subtraction) on the mass ranges
selected. Enter the row identification (A,B) to display the curves in a simple way.
6.
Select the
Range Limit
dialog.
Summary of Contents for FLEXAR SQ 300 MS
Page 1: ... 520 5 5 64 06 86 5 6 8 Chromera Chromatography Data System ...
Page 2: ......
Page 3: ...Chromera and Flexar SQ 300 MS User s Guide ...
Page 8: ...6 Flexar SQ 300 MS User s Guide ...
Page 9: ...Starting ...
Page 14: ...12 Flexar SQ 300 MS User s Guide ...
Page 15: ...Starting Chromera ...
Page 36: ...34 Flexar SQ 300 MS User s Guide ...
Page 37: ...Initial Process to Configure an Optimal Tune and Method on the SQ 300 MS Detector ...
Page 55: ...Creating Methods and Sequences ...
Page 68: ...66 Flexar SQ 300 MS User s Guide ...
Page 69: ...Starting Data Acquisition ...
Page 77: ...Analyze Results in Post Run ...
Page 104: ...102 Flexar SQ 300 MS User s Guide ...
Page 105: ...Evaluating Mass Spectra ...
Page 119: ......