128 . MultiScope System Microscope User's Reference
Aligning the MCT Detector
The detector must be aligned when the microscope is first installed in your system.
After that, it needs realignment only when it has been removed from the system
and renewed, or when you suspect that it is not receiving the maximum signal. You
will need the cross-head screwdriver that was supplied with the microscope.
WARNING
Laser radiation will become accessible if the MCT detector is
removed.
If the MCT detector is to be removed from a Spectrum BX or
Model 1600, first block off the laser radiation by selecting the
FT-IR’s internal beam path.
1.
Switch off the microscope and disconnect the mains supply.
2.
Remove the two screws that secure the microscope top cover; one is on the
right at the rear, and the other at the front underneath the MCT detector
cover. Lift off the cover vertically.
The red MCT dewar and the metal frame and springs that support it become
visible (Figure 43).
Figure 43 Interior of the Detector Area of the MultiScope System
Microscope
Summary of Contents for MultiScope System
Page 1: ...MultiScope System Microscope User s Reference ...
Page 5: ...Introduction ...
Page 14: ...14 MultiScope System Microscope User s Reference ...
Page 15: ...Warnings and Safety Information ...
Page 31: ...Overview of the MultiScope System Microscope ...
Page 44: ...44 MultiScope System Microscope User s Reference ...
Page 45: ...Getting Ready to Use the Microscope ...
Page 53: ...Tutorial Using the Microscope ...
Page 63: ...Preparing Samples ...
Page 83: ...Collecting Spectra with the Microscope ...
Page 95: ...Operating the Optional Equipment ...
Page 109: ...Applications ...
Page 122: ...122 MultiScope System Microscope User s Reference ...
Page 123: ...Maintenance ...