134 . MultiScope System Microscope User's Reference
MultiScope System Microscope Connected to a
Spectrum One or Spectrum 100 Series Spectrometer
The cable from the microscope to the spectrometer comes from the preamp inside
the microscope, to a connector marked
on the rear of the FT-IR.
Figure 45 MultiScope System Microscope Connected to the Spectrometer
The following table shows all of the connections from the pre-amp in the
microscope to the spectrometer:
Pl 1 (PCB)
D-type Connector
Line
Power
Requirement
1
11
Output signal
2
10
Output signal (0 V)
3
9
0 V
4
12
0 V
5
13
+12 V
250 mA
6
14
-12 V
200 mA