Collecting Spectra with the Microscope
.
85
Positioning the Sample
When you use the microscope to collect spectra, you must center the sample on
the sample stage before you collect the background spectrum. This enables you to
set the correct aperture for the sample and then use that aperture when collecting
the background spectrum.
1.
Take out any aperture that is in the retaining well.
2.
Place the sample on the sample stage.
If necessary, secure it with the slide clip:
Figure 18 Parts of the Sample Stage
3.
To set the microscope for viewing a sample, turn the View/IR knob to View.
The detector mirror moves out of the optical path, and the illuminator is
switched on.
4.
Set the Trans/Refl knob to the required position for viewing.
5.
Turn the Illuminator knob to adjust the light to an adequate level for viewing.
NOTE: If you need to increase the illumination during the procedures that follow,
move the knob to a higher setting.
Stage Controls
Slide Clip
Summary of Contents for MultiScope System
Page 1: ...MultiScope System Microscope User s Reference ...
Page 5: ...Introduction ...
Page 14: ...14 MultiScope System Microscope User s Reference ...
Page 15: ...Warnings and Safety Information ...
Page 31: ...Overview of the MultiScope System Microscope ...
Page 44: ...44 MultiScope System Microscope User s Reference ...
Page 45: ...Getting Ready to Use the Microscope ...
Page 53: ...Tutorial Using the Microscope ...
Page 63: ...Preparing Samples ...
Page 83: ...Collecting Spectra with the Microscope ...
Page 95: ...Operating the Optional Equipment ...
Page 109: ...Applications ...
Page 122: ...122 MultiScope System Microscope User s Reference ...
Page 123: ...Maintenance ...