108 . MultiScope System Microscope User's Reference
When all of these parts are in place, the gas entering through the inlet displaces air
from the path of the infrared beam and the sample area.
Purging the system
1.
Make sure that all parts of the purge system, as listed above, are in place.
2.
Set up the microscope.
3.
Place the sample in position.
4.
Purge the system for 15 to 20 minutes at a rate of approximately 10 L min
-1
.
5.
Collect the background spectrum and the spectrum of the sample.