Chapter 1. Programming Site Data
26
DigitalFlow™ XMT868i Programming Manual
1.4.2
Setting Up Inputs and Outputs
Set up the XMT868i’s inputs and outputs using the
I/O
submenu. While following the programming instructions, refer
to Figure 31 on page 10 and Figure 32 on page 11. Remember to record all programmed data in Appendix B,
Data
Records
.The
I/O
submenu consists of the following options:
•
Error Handling
- program the meter’s response during an error condition
•
Options
- set up any option cards and the
Slot 0
analog outputs
•
LCD
- set up the LCD display (refer to Chapter 2,
Displaying Data
)
To enter the
I/O
submenu:
1.
From the
PROGRAM
menu, scroll to
GLOBL
and press
[Enter]
.
2.
Scroll to
I/O
and press
[Enter]
.
Note:
In this section,
Slot 1
appears as an option only if a suitable option card is installed in Slot 1.
1.4.2.1
Setting Up Error Handling
This menu option lets you set how the XMT868i will handle the outputs for measurements and average (two-path)
measurements during an error condition. See Chapter 2,
Error Codes
, in the
Service Manual
for a discussion of the
built-in error codes.
1.
In the
Global I/O
option, scroll to
ERROR
and press
[Enter]
.
2.
Scroll to the desired
Error Handling
selection (as shown in Table 15 and Table 16 on page 27).
a.
If you selected
Other
, enter the
4-20 mA Error Level
(the number of milliamps the analog output will put out during
a fault condition). Enter a whole number between 0 and 22, and press
[Enter]
.
See Table 15 and Table 16 on page 27 for a description of the error handling options available and how the totalizers
and display respond to them for a single and two-channel meter.
Table 15: Error Options and Responses, Single-Channel Meter
Option
Output Response
Totalizer Response
HOLD
Holds the last “good” reading
Holds the last “good” reading and continues to
totalize, based on that reading
LOW
Forces the outputs to the low set point
Stops totalizing
HIGH
Forces the outputs to the high set point
Stops totalizing
HHIGH
Forces the outputs
10% above the high set point
Stops totalizing
OTHER
Allows user to enter 4-20 mA error level in mA.
Sends analog outputs to entered mA level when
a fault occurs.
Summary of Contents for DigitalFlow XMT868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 50: ...Chapter 1 Programming Site Data 42 DigitalFlow XMT868i Programming Manual...
Page 74: ...Chapter 4 Printing Data 2 DigitalFlow XMT868i Programming Manual...
Page 78: ...Chapter 5 Clearing Data 4 DigitalFlow XMT868i Programming Manual...
Page 79: ...DigitalFlow XMT868i Programming Manual 5 Appendix A Menu Maps Appendix A Menu Maps...
Page 80: ...Appendix A Menu Maps 6 DigitalFlow XMT868i Programming Manual...
Page 86: ...Appendix A Menu Maps DigitalFlow XMT868i Programming Manual...
Page 144: ...Appendix D PanaView Menu Maps for the XMT868i 70 DigitalFlow XMT868i Programming Manual...
Page 150: ...Appendix A PanaView Menu Maps DigitalFlow XMT868i Programming Manual 76...
Page 162: ...Warranty 6 AquaTrans AT868 User s Manual no content intended for this page...
Page 163: ......