CGA 351 User’s Manual
51
Chapter 4.
4.1.9.3
Setting the Span Setpoint
Enter the analog output setup menu, as described on page 47, and proceed as follows:
Note:
The
[Test], [Zero Trim]
and
[Span Trim]
options are also available in the above list box.
At the next prompt, enter the desired span setpoint.
The meter now automatically returns to the “Analog Output” prompt. Either press
[ESC]
until you exit the
User
Program or proceed to the appropriate section to continue setting up the analog output.
Table 82:
User Aout
Use the [
] and [
] keys to
select
[Span Setpoint]
, and
press
[ENTER]
.
Zero Setpoint
Span Setpoint
Test
POWER
ESC
ENTER
Table 83:
User Aout Span-Set
Use [
] and [
] to position
the cursor under the desired
character. Then, use [
] and
[
] to increment the value.
Press
[ENTER]
when done.
mV
0000.00
p
POWER
ESC
ENTER