CGA 351 User’s Manual
29
Chapter 4.
•
Trip Type:
specifies whether the alarm relay triggers when the live measurement value moves
above or below the
trip value.
•
Trip Value:
specifies the live measurement value at which the alarm relay triggers.
•
Deadband Value:
specifies a small range of live measurement values, centered on the trip value, for which the
alarm relay does not trigger or reset. This prevents the alarm relay from repeatedly triggering and resetting when
the live measurement value is very close to the trip value.
•
Test:
permits the manual triggering and resetting of the alarm relay, to verify that it is working properly.
•
Fail-Safe:
specifies whether the alarm relay is energized during normal operation or during an error condition.
Go to the appropriate sub-section to set the desired alarm feature.
Note:
The setup of Alarm-1 is used as an example in this manual. To set up Alarm-2, -3 or -4, return to the previous
prompt and select the
desired
option. Then, repeat the following steps.
4.1.4.1
Setting the Trip Type
Enter the alarm setup menu, as described on page 28, and proceed as follows:
Note:
The
[Test]
and
[Fail-Safe]
options are also available in the above list box.
Setting the Trip Type (cont.)
Note:
The
[Below with Fault]
option is also available in the above list box.
The meter now automatically returns to the “Alarm-Features” prompt. Either press
[ESC]
until you exit the
User
Program or proceed to the appropriate section to continue setting up the alarms.
4.1.4.2
Setting the Trip Value
Enter the alarm setup menu (see page 28) and proceed as follows:
Table 27:
Alarms..Alarm-Feat
Use the [
] and [
] keys to
select
[Trip Type]
, and press
[ENTER]
.
Trip Type
Trip Value
Deadband Value
POWER
ESC
ENTER
Table 28:
Alarms A..Trip-Type
Use the [
] and [
] keys to
select the desired trip type,
and press
[ENTER]
.
Above
Below
Above with Fault
POWER
ESC
ENTER
Summary of Contents for CGA 351
Page 2: ......
Page 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Page 32: ...CGA 351 User s Manual 26...
Page 94: ...CGA 351 User s Manual 88...
Page 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Page 104: ...Appendix B 98 CGA 351 User s Manual...
Page 112: ...CGA 351 User s Manual 106...
Page 118: ......