Chapter 4.
32
CGA 351 User’s Manual
4.1.4.4
Testing the Alarm Relays
Enter the alarm setup menu, as described on page 28. To verify that the alarm relay is functioning properly, the alarm
may be operated manually as follows:
Note:
The
[Trip Type]
and
[Fail-Safe]
options are also available in the above list box.
The following options are available in the [
TEST]
submenu:
•
Normal:
sets the alarm relay to its proper state, as dictated by the current live measurment value.
IMPORTANT:
Always return the alarm relay to this state at the conclusion of the testing.
•
Trip:
manually triggers the alarm relay to its “ON” state.
•
Reset:
manually returns the alarm relay to its “OFF” state.
The meter now automatically returns to the “Alarm-Features” prompt. Either press
[ESC]
until you exit the
User
Program or proceed to the appropriate section to continue setting up the alarms.
Table 33:
Alarms..Alarm-Feat
Use the [
] and [
] keys to
select
[Test]
, and press
[ENTER]
.
Trip Value
Deadband Value
Test
POWER
ESC
ENTER
Table 34:
Alarms..Alarm-Test
Use the [
] and [
] keys to
select the desired
TEST
option.
Press
[ENTER]
when done.
Normal
Trip
Reset
POWER
ESC
ENTER
Summary of Contents for CGA 351
Page 2: ......
Page 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Page 32: ...CGA 351 User s Manual 26...
Page 94: ...CGA 351 User s Manual 88...
Page 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Page 104: ...Appendix B 98 CGA 351 User s Manual...
Page 112: ...CGA 351 User s Manual 106...
Page 118: ......