Chapter 2.
8
CGA 351 User’s Manual
2.1.2
Choosing an Installation Site
All environmental and installation factors should have been discussed with a Panametrics Sensing applications
engineer or field sales person at the time the CGA 351 was ordered. Thus, the equipment should be suited to the
application and the planned installation site. Before installing the unit, however, read the following guidelines to verify
that the best installation site has been chosen for optimum instrument accuracy and reliability. Make sure that:
•
the CGA 351 enclosures are suitable for the environmental conditions at the installation site.
•
the analyzer system will be installed in a location with little or no vibration.
•
the ambient temperature at the installation site is between –25° and 50°C (–13° and 122°F).
•
there are no restrictions in the discharge line (i.e. flow control valves downstream of the meter).
•
the CGA 351 system is installed in a location that provides ready access for programming, testing, and servicing
the unit.
•
the cables will be protected from excessive physical strain (bending, pulling, twisting, etc.). Do not subject the
cables to temperatures above 65°C (149°F) or below –50°C (–58°F).
•
the line voltage used at the planned installation site corresponds to the factory preset line voltage rating for the
meter.
Summary of Contents for CGA 351
Page 2: ......
Page 4: ...Warranty 4 CGA 351 User s Manual no content intended for this page...
Page 32: ...CGA 351 User s Manual 26...
Page 94: ...CGA 351 User s Manual 88...
Page 98: ...Chapter 6 92 CGA 351 User s Manual no content intended for this page...
Page 104: ...Appendix B 98 CGA 351 User s Manual...
Page 112: ...CGA 351 User s Manual 106...
Page 118: ......