26
6.3.3 Replacing the batteries
NOTICE
Risk of battery damage.
Ensure correct polarity is used when connecting the batteries. The black ca-
ble must be connected to the negative pole and the red cable to the positive pole.
Preparation:
❒
Turn the control unit off.
❒
Remove the fuse (see Section 6.2).
❒
Remove the battery cover (see Section 6.3.2).
Tools:
❒
Open-end or combination wrench, size 11
Steps:
1. Unlock the release bolt and remove the track.
fig. 12 Unlocking the battery drawer and removing the track
2. Disconnect all battery cables.
3. Lift out the batteries by the handles and replace if required.
fig. 13 Removing the batteries by the handles