105
Technical data for the first actuator output:
Voltage
0…100 % V
Batt
Current for 20 seconds
20 A
Steady current
10 A
Technical data for the second actuator output:
Voltage
0…100 % V
Batt
Current for 20 seconds
15 A
Steady current
7 A
The actuator outputs are short-circuit protected. However, short-circuiting the actuator output
against the negative pole of the battery while the actuator is in operation may damage the out-
put. The actuators have high impedance when they are not in operation.
8.3 Seat module
The controller is attached to the wheelchair with two screws and can be aligned in any position
(see Section 6.16.6).
1
2
3
4
5 6 7
fig. 73 Seat module connection schemata
1 IN-A, IN-B, IN-C, IN-D, IN-E connections
2 RSW
3 A1 - A5 connections (actuators A1 - A5)
4 BUS connection
5 BUS connection
6 BATT bridge / external supply connection
7 12 V DC (connection must be activated by parameter -> On/Off!)