Chapter 8 - Technical Specification
Immunity test
IEC60601 compliance
test level
Electromagnetic environment-guidance
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power supply
lines
±1 kV for input / output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC61000-4-5
±1 kV line to line
±2 kV line to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
lines
IEC61000-4-11
<5% U
T
(>95%dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
device requires continued operation during mains
interruptions, it is recommended that the device be
powered from an uninterruptible power supply.
<5% U
T
(>95% dip in U
T
)
for 5 seconds
If the user of the device requires continued operation
during mains interruptions, it is recommended that the
device be powered from an uninterruptible power
supply.
Power frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to the application of the test level.
Conductive RF
IEC61000-4-6
3Vrms
150kHz to 80MHz
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2√P
Part Number: G110230-003-ENG
Page 75 of 84