Chapter 8 - Technical Specification
9 Technical Specification
Your device is a medical device and, as such, should be operated within the technical parameters and
instructions defined in this User Guide. Read the warnings and cautions before using your device (see
If you have any questions regarding the correct use of your device, please contact your Optos
representative (see
9.1 Safety designation (IEC 60601-1)
In order to provide reasonable interference protection against radio communication services, this device
has been tested and complies with 47 CFR Part 15, Sub Part B of the FCC (U.S.A.) rules and IEC/EN
60601-1-2 standard requirements at the premises of a typical user, and the CISPR11 standard emission
limits (CISPR11 Group 1 Class B) (see
Guidance and manufacturer declaration – electromagnetic
Operation is subject to the following two conditions:
1. This device may
NOT
cause harmful interference.
2. This device
MUST
accept any interference received, including interference that may cause undesired
operation.
9.1.1 Scan head
TABLE 10: Scan head safety designation details
Feature description
Compliance
Type of protection against electrical shock
Class 1
Degree of protection against electrical shock
Type B applied part
Type B - relates to the allowable maximum leakage
current which can flow from the applied part. The limits
for this are defined in IEC/EN 60601.
Applied parts
Forehead pad, chin pad and face pad
Degree of electrical connection between
equipment and patient
The equipment is designed as a non-electrical
connection to the patient.
Degree of mobility
Transportable
Mode of operation
Continuous
Device type
Medical
Supply connection
100 to 240Vac, 50/60Hz, (IEC60601-1)
Electrical Protection
T5A H250V (5 mm x 20 mm)
Part Number: G110230-003-ENG
Page 69 of 84
Copyright 2019-2021, Optos plc. All rights reserved.