Chapter 1 - Warnings and cautions
CAUTION
Unauthorized changes or modifications will invalidate the warranty of this device.
The manufacturer is
NOT
responsible for any interference caused by using:
l
Interconnect cables
NOT
supplied by Optos
l
Unauthorized changes or modifications to this equipment
Do
NOT
use devices which intentionally transmit RF signals (cellular phones,
transceivers or radio controlled devices) in the vicinity of this equipment, as it may cause
performance outside the published specifications (see
Adhering to the distance separation recommended will reduce the disturbances
recorded at the image level, but may
NOT
eliminate all disturbances (see
recommended separation distances
on page 76). However, when installed and
operated as specified herein the system will maintain its essential performance by
continuing to acquire, display and store diagnostic quality images safely.
It is recommended that the image server is not used adjacent to or stacked with other
equipment.
The use of parts and cables other than those specified may result in degraded
electromagnetic compatibility of the system.
1.6 Device failure
WARNING
In the unlikely event you hear a loud noise from the device, see smoke or smell burning,
stop imaging and isolate the device by unplugging from the mains power supply.
If you suspect the device may be faulty (see
If the fault still persists, please contact your Optos representative (see
CAUTION
The device is isolated from the mains by the main power cable and IEC connector. In an
emergency, the device should be isolated by unplugging from the mains power supply.
Ensure the mains socket is easily accessible.
1.7 Cleaning guidelines
The device cleaning instructions are detailed in this User Guide (see
on page 60). Adhere to the
following cautions while cleaning the device.
CAUTION
Cleaning patient contact parts
In order to prevent any contamination always clean and disinfect parts of the equipment
which will be in contact with the patient (see
Adequate cleaning and disinfection is necessary to prevent disease transmission.
Thoroughly clean and disinfect the equipment surfaces that contact the patient and all
equipment surfaces likely to become soiled during use.
Do
NOT
attempt to wash or immerse the patient contact parts (for example, forehead
pad, chin pad and face pad) in water or cleaning fluid.
Page 16 of 84