- 46 -
❖
Focus the image, the characteristic point move away its original position along another
direction, adjust another knob of final aperture holder, and make this point move back to its
original position. Repeat this step, until this point almost stay in the original position when
focusing or defocusing.
5.
Adjust the magnification to 1000X-2000X. Repeat step 3 and 4.
Note:
If you want to observe the image at high magnification, adjust the magnification over
5000X, repeat the step 3 and 4 again.
6.
Any changes of final aperture, will break the Astigmatism status. Therefore, when finished Final
Aperture Alignment, the user must do astigmatism correction. (About astigmatism correction,
please refer to 5.4)
5.3.3.
Final Aperture Selection
1.
Three apertures have been installed in the final aperture holder.
Aperture Size:
Anode aperture - 300um
Aperture of condenser lens- 300 um
Final aperture (optional) - 300um, 200um and 100um (6900: 70um, 50um or 30um)