- 26 -
3.4.13.
Stigmator
❖
Coil of stigmator is installed in scanning coil assembly. The function of stigmator is to correct
elliptic electron spot caused by aperture pollution or other distortions of lens magnetic field.
The stigmator can correct astigmatism rapidly and easily.
❖
Clicking the [Stigmator] button
on panel can start stigmator adjustment. The Stigmator
control contains a box area, two scroll bars and a display label.
❖
Dragging horizontal and vertical scroll bar of [Stigmator] can control stigmator. Dragging
leftward or rightward can change stigmator control value in X direction, while dragging
upward and downward can change stigmator control value in Y direction.
❖
Directly clicking in the box area or dragging the cross line in the box can adjust two directions
simultaneously. Right clicking in the box area can return control value of XY to zero quickly.
❖
For users
’
convenience, adjusting stigmator can operate as follows: move cursor on the
collected image, press right and drag.
3.4.14. Scan Rotation
❖
The Scan (Raster) Rotation function electrically rotates the scan raster or scan line from 0 to
360 degree.
❖
Clicking the [Scan Rotation] button
on panel can start the rotation adjustment.
❖
Dragging knob of [Scan Rotation] clockwise/anticlockwise with mouse can control it from 0
to 360 degree.