- 50 -
Note:
WD is the distance when the image is clearest. After adjust the movement of Z direction,
focus the image. When the image is clearest, the display of WD is the real value.
❖
Rotation of specimen stage: Rotate the knob R in the stage, the specimen holder can rotate
at any degree from 0 to 360
❖
Specimen tilt: Rotate the knob T in the stage, the specimen holder can be tilted from -5°to
+90°.
5.5.3.
Specimen Grounding Alarm
❖
A warning device has been installed in the stage. When the specimen touch the bottom pole
piece, a buzzer will alarm. At this moment, do not adjust the knob X, Y, T of specimen stage.
Only rotate the knob Z clockwise to increase the WD.