5.4 LIGHT EMISSION
Keratron™ Nova contains within it different light sources, it is a
LED Class 1 device
in accordance with IEC60825-1.
In all operating conditions
the maximum output power is lower than that power
specified in the table below:
Operating condition
Wavelength
-
λ
peak
-
Emitted power
-
P
meas
-
Topography (visible LED)
594 nm
140μW (<LEA Class 1)
Pupillometry (Infrared LED)
950 nm
150μW (<LEA Class 1)
If the cone is removed while capturing or calibrating, the lamp is turned off in order
not to expose the operator to dangerous radiation.
5.5 CIRCUIT DIAGRAMS
Upon request, OPTIKON 2000 SpA provides wiring diagrams, parts lists,
descriptions, calibration instructions or other information to support the operator’s
qualified personnel in case of repair of the serviceable elements of the unit.