5.3 ELECTROMAGNETIC COMPATIBILITY TABLES
5.3.1 GUIDE AND MANUFACTURER’S
STATEMENT-ELECTROMAGNETIC
EMISSIONS
The Keratron™ Nova product is designed to work in the electromagnetic environment
specified below. The customer or user of Keratron™ Nova must guarantee that it is used
only in such an environment.
Emission test
Conformance
Electromagnetic environment - guide
RF emissions
CISPR 11
Group 1:
The Keratron™ Nova uses RF energy
only for its internal workings. Therefore,
its RF emissions are very low and not
likely to cause any interference in the
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Keratron™ Nova is suitable for use
in all areas, including domestic and
those directly connected to public low-
voltage power supply that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Emissions of voltage
fluctuations/flicker
IEC 61000-3-3
Conforming