59
Figure 54: Datapoint changing its position after the tip is getting in contact with a surface: probe out of contact (a), noise
caused by probe moving during automated find surface procedure or by manually moving the Z-stage (b), probe in
contact with surface resulting in datapoint position change (c).
The ‘Settings’ menu allows to set all current parameters as default values. When restarting the
OP1550 interferometer the last saved values will be loaded.
The Laser, ADC, Output, Modulation and Maintenance menu
The right side of the screen is reserved for the Laser, ADC, Output, Modulation and Maintenance
menu. To scroll through the menus, move the cursor to the section right of the graph area, and
press the left or right arrow.
Figure 55: Laser (a), ADC (b), Output (c), Modulation (d) and Maintenance menu (e) of the OP1550 interferometer.
The ‘Laser’ menu displays the laser controls. Always operate channel 86 of the laser: this
corresponds to 1530nm, which is also set in the maintenance menu of the software.