54
optics11.com
In the probe configuration menu, a tick-
box ‘Store Load
-
Displacement graph as BMP’
can be ticked
to store each load-displacement curve as BMP in addition to the tab-separated text file containing
the raw data.
Options menu
The ‘Options menu’ allows
setting of a
number of variables that define the ‘Find
-
surface’ function,
matrix scan settings, and model selection and parameters (Figure 50).
The ‘dZ at XY move’ field controls the transportation height of the probe when moving to a new
coordinate in a matrix scan, the ‘Z above surf.’ field controls the offset in Z
-direction of the probe tip
to the sample surface after detecting the surface using the automated find-
surface function, the ‘dX
before scan’ field controls the offset in X
-direction of the probe tip with respect to the surface after
the surface has been determined.
Figure 50: Options menu.
The Model settings section, displaying the ‘Oliver & Pharr’ and ‘Her
t
zian contact’ buttons, allow one
to select the model of choice to be applied for the indentation or matrix scan. In case the Oliver &
Pharr model is selected, the ‘Pmax
%’ and ‘Pmin
%’ fields control the section of the unloading curve,
expressed in a percentage of the maximum load in the unloading part of the curve, that is used to
make the slope fit. In case the Hert
zian model is selected, the ‘% of Pmax’ field controls the part of
the curve to be fitted, expressed in a percentage of the maximum load in the loading part of the
Summary of Contents for CHIARO NANOINDENTER