curve. One can also choose to not apply a model by clicking whichever model is highlighted a
second time, which will grey out both sets of model parameters.
Please note only the Young’s Modulus of the selected model is currently saved.
Under the ‘Other’ section of the options menu you can set a calibration depth, which refers to the
indentation depth while performing a geometrical calibration. It is recommended to set this value to
a similar indentation depth, which is accomplished during a standard measurement.
Maintenance menu
The maintenance panel is protected by a password, as any change in these settings may
compromise the performance of the Piuma Nanoindenter.
Please note that changing parameters in the maintenance menu can cause unexpected behavior
and potentially break the hardware on the Piuma. Use caution when changing parameters.
It is recommended to only use the maintenance menu when switching between the Piuma and
Piuma
indenter head for XYZ stage calibration (‘XYZ’ tab) or when adjusting the DAQ rate. The rate
can
be set in the ‘DAQ & Calib’ tab in the Maintenance menu to 1Hz up to 16000Hz or 1Hz to
12000Hz, if a heater is being used.
Figure 51: DAQ rate adjustment.