70
Maintenance
Setting
On the sensor selection display, select a
sensor to set using
, and press
.
Maintenance timer setting must be set for each sensor.
If [ALL] is selected, the setting is not available.
Select [Maintenance Timer] using
,
and press
.
1
2
0DLQWHQDQFH
6HQVRU6HOHFW
㸯
S+ࠉ6(S+'
㸰
7XUE7&'
㸱
'2'26
㸲
ࠉ
$//
Output
item
Sensor
model
Sensor channel
0DLQWHQDQFH
㸱
'2'26
+2/'
+2/'7LPHUࠉࠉPLQ
0DLQWHQDQFH7LPHU2))
6HQV&DS5HSODFHPHQWGD\V
u
Maintenance Timer
A maintenance timer can be set to an arbitrary cycle and notified when the set cycle has
elapsed.
This is used for notification of sensor maintenance, replacement of consumables, and
overhaul timing.
When the maintenance timer has expired, "MAINT" is
displayed on the measurement value display of the sensor.
Notifications can be checked by the relay output
("Maintenance output setting" P.46) and the cloud server (when connected to the GW).
For more information about checking notifications on the cloud server, refer to the cloud
server’s help.
Output Item
Setting range
Factory default setting
Maintenance
Timer
OFF
7 to 1095 days (by 1 day)
OFF
●
Setting maintenance timer
Set the maintenance timer.
The setting range is shown below.
㸯
0$,17
'2
5/<
PJ/
Υ