86
Error details
Cause
Action
Data cannot be
sent from the
gateway.
The send channel
setting is not complete.
Output to the gateway is not available unless
transmission item to the transmission channel as
well as the transmission interval have been set.
Refer to "Send channel setting" P.28 to configure
the setting.
An error occurred in the
gateway.
Follow the description of the gateway's instruction
manual.
The cloud settings are
wrong.
Follow the Cloud Help for confirmation.
4-20 mA output
is not available.
4-20 mA output setting
has not be finished yet.
Refer to "5.3 4-20 mA Setting" P.30 to configure
the setting.
Connector for SC-U-EB
is not properly connected.
Turn off power, check the connector, and refer to
"Sensor Input/Output Expansion Board SC-U-EB
Instruction Manual", "4. Installation" to wire the
cable properly.
The cable is not properly
wired.
Turn off power, check the wiring, and be sure to
wire properly.
The cable is broken.
Make sure that the cable is not damaged.
If any damage is found, contact the dealer.
The transmitter is out of
order.
The transmitter must be repaired. Contact the
dealer.
Three or more
4-20 mA output
channels cannot
be connected.
SC-U-EB is not connected. To connect two or more 4-20 mA output channels,
SC-U-EB must be connected to the transmitter.
Then up to four 4-20 mA output channels can be
connected.
The 4-20 mA
output is
incorrect.
The measurement value
is held.
When
+2/'
is displayed in the sensor status, the
measurement value is retained.
When the time set by the hold timer elapses, the
setting is canceled.
To cancel manually, refer to “Canceling
measurement value hold” P.68 to cancel the
hold.
It is included in each
menu.
If it is included in each menu, the measurement
value is retained.
When the measurement value display is returned,
the hold of the measurement value is canceled.