13
No.
Name
Description of Functions
(1) HOME
The indication returns from menu to measurement value display.
Selection button
(2)
UP/DOWN/
LEFT/RIGHT
• UP/DOWN:
Used to select an item in the setting menu.
Or, used to increase/decrease a value of the selected digit in the
numerical input.
• LEFT/RIGHT:
Used to switch between pages or select an item in the setting
menu.
Or, used to select a digit in the numerical input.
In addition, pressing both left and right simultaneously while
entering a value resets the value to the default value.
(3)
ENTER
Used to determine the selected item in the setting menu.
Or, used to determine the entered value in the numerical input.
(4)
BACK
Used to cancel an item in the setting.
Or, used to return to the previous display if no item is being selected.
Function button
(5) CAL
Used to calibrate the sensor ("6 Calibration" P.60).
(6)
MAINT
Used to retain a measurement value "measurement value hold" or to
configure the hold timer setting ("8 Maintenance Setting" P.66).
(7)
CLEAN
Used to perform cleaning of the sensor or setting of cleaning time
("9 Cleaning" P.73).
(8)
SETTING
Used to perform settings such as input/output and date/time
("5 Setting Menu" P.20).
(9) ADJUST
Used to correct a measurement value ("7 Adjustment" P.63).
(1)
(4)
(5)
(6)
(7)
(9)
(8)
(2)
(3)
Operation button