Measurement Settings
78
ZG User’s Manual
3
Adjust the bottom right of the
desired measurement area by the
↑
,
↓
,
←
and
→
keys, and press the
SET key.
The measurement region is set to
the specified area.
Profile
Noise filtering, output at measurement failure, and other options can be set in more detail.
Adjust the conditions when the measurement cannot be performed properly.
Setting of profiles is enabled only in the EXP mode.
Setting the average number of times
Changes in data are smoothed out using the average values of adjacent data. Smoothing
is performed in the “width direction (X-axis direction).”
FUN mode-[MEAS]-[IMAGE]-[PROFILE]-[AVERAGE]
Setting value
Description
1, 2, 4, 8, 16, 32, 64
Sets the number of data to average.
(default value: 4 when the CCD mode is NORMAL or HI-RESO, and
1 when the CCD mode is HI-SPEED)
↑↓
10
.
000mm
←→
4
50
LINE
S
ET
M
E
AS
/
C
H
A
N
G
E
M
E
AS
/
R
E
G
I
O
N
1C
H
A
N
G
E
[1]
OK
MEAS/AVERAGE
[64]
OK
MEAS/AVERAGE
Summary of Contents for ZG -
Page 12: ...10 ZG User s Manual MEMO ...
Page 44: ...Overview of Settings and Measurement 42 ZG User s Manual MEMO ...
Page 60: ...Functions Operations Used during Operation RUN Mode 58 ZG User s Manual MEMO ...
Page 110: ...System Settings 108 ZG User s Manual MEMO ...
Page 156: ...Serial Communication 154 ZG User s Manual MEMO ...
Page 206: ...204 ZG User s Manual MEMO ...
Page 207: ...205 ZG User s Manual 5 APPENDICES MEMO ...
Page 208: ...206 ZG User s Manual MEMO ...