![background image](http://html1.mh-extra.com/html/omron/k3hb/k3hb_technical-communications-users-manual_743344042.webp)
1-21
Commands and Responses (CompoWay/F Communications)
Section 1-9
Co
m
p
o
W
a
y
/F
Co
m
m
u
n
ica
tio
n
s
Response
Response codes: The above indicates a normal end. For the response codes,
see
1-7 Operation Commands (CompoWay/F Communications)
1-9-18 Read Machine Attributes
Command
This command reads the K3HB model and communications buffer size.
The command can be used in any state of the K3HB.
Response
Response codes: The above indicates a normal end. For the response codes,
see
1-7 Operation Commands (CompoWay/F Communications)
Model
1-9-19 Controller Status Read (CompoWay/F Communications)
Command
This command reads the operation state of the K3HB.
The command can be used in any state of the K3HB.
Response
"30"
"05"
"0000"
Response code
SRC
MRC
"05"
"03"
SRC
MRC
A
B
Symbol
Series
Symbol
Input
X
Analog input
VD
DC voltage
AD
AC voltage
VA
DC current
AA
AC current
V
mV input
LC
mV
S
Linear Sensor input
SD
Linear Sensor
H
Temperature input
TA
Temperature
R
Rotary pulse input
NB
NPN/voltage pulse input
PB
PNP input
P
Time interval input
NB
NPN/voltage pulse input
PB
PNP input
C
Up/Down counting pulse
input
NB
NPN/voltage pulse input
PB
PNP input
"05"
"03"
"0000"
"00D9"
Buffer size
Format
Response code
SRC
MRC
B
-
K
∗
Bytes 8 and 9 are blank
3
H
"06"
"01"
SRC
MRC
"06"
"01"
"0000"
Related information
Operation
state
MRC
SRC
Response code
Summary of Contents for K3HB
Page 2: ...K3HB Digital Indicators Communications User s Manual Revised March 2005...
Page 3: ...iv...
Page 15: ...xvi...
Page 19: ...xx...
Page 21: ...Part 1 CompoWay F...
Page 69: ...Part 2 DeviceNet...
Page 95: ...6 10 Ladder Programming Examples Section 4 3 Remote I O Com munications...
Page 111: ...8 8 Message Communications Performance Section 6 2 Communications Performance...
Page 127: ...Part 3 BCD Output...
Page 147: ......